12-9
Troubleshooting
Operator Manual
129373-496
Alarm
Description
Screen with Operator Instructions
12.2.11 Reference
Deviation Alarm
Occurs if the two
temperature sensing
elements in the cham-
ber drain probe read
more than 1°C apart.
Screen with Operator Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
249
REFERENCE DEVIATION
CHAMBER:
00.0 C
0.0 psig
RECORDER: 60.0 C
STERILIZER WILL:
•
PREVENT NEW CYCLE FROM BEING
STARTED UNTIL ALARM IS CLEARED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
SERVICE INFORMATION:
244
EXHAUST RATE TOO SLOW
→
→
→
→
→
CHAMBER EXHAUSTED SLOWER
THAN THE EXPECTED RATE.
CAUSES AND CORRECTION:
1. SOLENDOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S40
2. SOLENMOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S03
3. CHAMBER STEAM TRAP MALFUNCTION
→
→
→
→
→
REPAIR
4. CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
CLEAR
EXIT
FEED
ALARM
Occurs if liquid cycle
slow exhaust rate is too
slow.
12.2.10 Exhaust Rate Too
Slow
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
243
EXHAUST RATE TOO SLOW
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING
TO OPTIMAL COOLING RATE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Screen with Service Instructions