![Amsco Century Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190079.webp)
8-7
Programming Cycle Values
Operating Instructions
129373-496
•
Pressing
MAX. PRESS.
advances display to screen #31. Enter maximum
pressure value using the numeric keypad. Allowable maximum pressure
range is 0-45 psig. Pressure value appears on display as it is being
entered.
NOTE: If an out-of-range cycle value is entered, a display screen will
appear indicating the incorrect value and the allowable range. Display
screen will automatically return to previous screen, allowing operator to
enter an in-range cycle value.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#20.
31
CHANGE VALUES - MAX. PRESS.
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET MAX. PRESS.
26.0 psig
The max. press. is
the pressure level
that the charge
pulses will reach.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#20.
•
Pressing
MIN. PRESS.
advances display to screen #30. Enter minimum
pressure value using the numeric keypad. Allowable minimum pressure
range is 0-29.9 inHg. Pressure value appears on display as it is being
entered.
NOTE: If an out-of-range cycle value is entered, a display screen will
appear indicating the incorrect value and the allowable range. Display
screen will automatically return to previous screen, allowing operator to
enter an in-range cycle value.
30
CHANGE VALUES - MIN. PRESS.
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET MIN. PRESS.
10.0 inHg
The min. press. is
the pressure level
that the vacuum
pulses will reach.