Troubleshooting
Operator Manual
129373-496
STATUS..
SERVICE INFORMATION:
270
ATMOSPHERIC PRESSURE ALARM
→
→
→
→
→
THE ATMOSPHERIC PRESSURE IS
GREATER THAN 1 PSI FROM THE
CALIBRATED ATMOSPHERIC PRESSURE
CAUSES AND CORRECTION:
1.
THE STERILIZER IS AT A HIGHER
ALTITUDE THAN WHERE IT WAS
ORIGINALLY CALIBRATED (FACTORY)
→
→
→
→
→
SET ALTITUDE
2.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3.
LOOSE CONNECTION IN TRANSDUCER
WIRING
→
→
→
→
→
REPAIR
4.
TRANSDUCER FAILED
→
→
→
→
→
REPLACE
→
→
→
→
→
RECALIBRATE