10-14
129373-496
Operating Instructions
Out of Cycle Options
10.8 Standby
The Standby option allows sterilizer to be manually switched from the Operating
mode to a Standby mode. When in Standby mode, all valves to the sterilizer
are off, including jacket steam and water.
NOTE: Sterilizer should be manually placed in the Standby mode if automatic
utility shutdown feature (Section 8) is not selected and the sterilizer will not be
used for an extended period of time (e.g., overnight).
1. Press
STANDBY
on screen #87.
87
OUT OF CYCLE OPTIONS
PREVIOUS
SUPER-
VISORY
DISPLAY
SERVICE
SENSORS
MODE
MAIN
MENU
STANDBY
2. All valves to the sterilizer are automatically shut off and display returns to
screen #0.
Pressing PREVIOUS returns display to
screen #13.
Pressing MAIN MENU returns display to
screen #1.
Pressing HEXAWAVE advances display
to screen #1, turns on all valves and
places sterilizer in the Operating mode.
0
S T E R I S S C I E N T I F I C
PREPARED FOR:
STERIS
SCIENTIFIC
CUSTOMER
STERILIZER ID:
VACO1