![Amsco Century Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190077.webp)
8-5
Programming Cycle Values
Operating Instructions
129373-496
4. Screen #20 is the first of three menu screens which list the programmable
values for the Prevacuum cycle type. To change any of the listed cycle
values, press the corresponding button.
NOTE: A definition of each cycle value is given on the corresponding display
screen.
Pressing CYCLE MENU returns display
to screen #15.
Pressing NEXT advances display to
screen #21.
Pressing MAIN MENU returns display to
screen #1.
20
CHANGE VALUES - 1. PREVAC (PREVAC)
CYCLE NAME = PREVAC
PURGE TIME = 1:00
PULSES = 4
MAX. PRESS. = 26.0 psig
MIN. PRESS. = 10.0 inHg
NEXT
MAIN
MENU
CYCLE
MENU
•
Pressing
CYCLE NAME
advances display to screen #77. This screen
allows operator to enter a custom name for the selected cycle.
Enter customized cycle name, maximum of eight characters, using the
alpha-numeric keypad. Cycle name appears on display as it is being
entered.
NOTE: Custom cycle name appears inside the corresponding touch
screen button on screens #2 and #15, along top of corresponding in-cycle
screens (#9) and on corresponding in-cycle printouts.
77
CYCLE NAME - PREVAC TYPE
RETURN
CYCLE NAME
---------------
←
←
←
←
←
→
→
→
→
→
Pressing
←
or
→
moves the cursor to the
left or right, respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#20.
Cycle Name
(Cycle Type)
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T
U
V
W
X
Y
Z
SPACE
0
1
2
3
4
5
6
7
8
9