12-14
129373-496
Operator Manual
Troubleshooting
STATUS..
SERVICE INFORMATION:
264
DISPLAY ERROR
→
→
→
→
→
FAILURE TO COMMUNICATE
TO THE DISPLAY
CAUSES AND CORRECTION:
1.
LOOSE DISPLAY CABLE CONNECTION
P1A TO DOOR
'A' AND/OR
P1B TO DOOR 'B'
→
→
→
→
→
REPAIR
→
→
→
→
→
REPLACE CABLE
2.
FAULTY DISPLAY BOARD ON DOOR 'A'
AND/OR DOOR 'B' SIDE
→
→
→
→
→
REPLACE
3.
FAULTY CONNECTION IN DISPLAY BOX
→
→
→
→
→
REPLACE
4.
FAULTY CONNECTION IN CONTROL BOX
→
→
→
→
→
REPLACE
PAPER
SERVICE
EXIT
FEED
MODE
12.3.5 Display Error
Occurs if the control
cannot communicate
with the display.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
263
DISPLAY ERROR
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Screen with Service Instructions
Screen with Operator Instructions
12.3.6 Atmospheric
Pressure Alarm
Occurs if atmospheric
pressure is greater than
(1 psi) from the cali-
brated atmospheric
pressure.
Alarm
Description
Screen with Operator Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
269
ATMOSPHERIC PRESSURE ALARM
CHAMBER:
000.0 C
0.0 psig
ALTITUDE SETTING: 0 TO 1000 FEET
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Occurs if the control
cannot communicate
with the display.