6-17
Sterilizer Operation
Operating Instructions
129373-496
• Retract Seal
– Steam is exhausted from the door seal.
9
GRAVITY CYCLE
PHASE:
RETRACT SEAL
DOOR WAITING TO BE UNSEALED
CHAMBER:
32.0 C
00.0
psig
PROJECTED CYCLE COMPLETION TIME:
00:20
m:s
STATUS
ABORT
NOTE: If sterilizer is equipped with double doors and interlock type 1 or 3 is
programmed for this cycle, operator must select which door to unseal by
pressing UNSEAL DOOR on the appropriate touch screen. Refer to S
ECTION
8 for description of interlock types.
• Complete
– The complete tone sounds and the cycle summary and end-
of-cycle messages are printed.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
7
GRAVITY CYCLE
PHASE:
COMPLETE
OPEN DOOR(S) TO UNLOAD
CHAMBER:
32.0 C
00.0 psig
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
7. Once cycle is complete, open chamber door and unload sterilizer. Display
returns to the main menu (screen #1).
WARNING – BURN HAZARD:
Sterilizer, rack/shelves, and
loading car will be hot after
cycle is run. Always wear pro-
tective gloves and apron (also
a face shield, if processing
liquids) when removing a pro-
cessed load. Protective gloves
and apron must be worn when
reloading sterilizer following
the previous operation.
WARNING – SLIPPING HAZ-
ARD: To prevent falls, keep
floors dry by immediately wip-
ing up any spilled liquids or
condensation in sterilizer load-
ing and unloading areas.