6-29
Sterilizer Operation
Operating Instructions
129373-496
6.8 Dart
®
Test Cycle
Operation
WARNING – STERILITY AS-
SURANCE HAZARD: Load ste-
rility may be compromised if
the biological air removal or air
leak test indicates a potential
problem. If these indicators
show a potential problem, refer
the situation to a qualified ser-
vice technician before using
the sterilizer further.
WARNING – PERSONAL IN-
JURY HAZARD: When closing
the chamber door, keep hands
and arms out of the door open-
ing and make sure opening is
clear of any obstructions.
The Dart (Bowie-Dick) Test cycle is designed to test and document the
adequacy of air removal from the chamber and a sample challenge load. Refer
to "Testing for Prevacuum Efficiency" in
S
ECTION
3, T
ECHNIQUES
OF
S
TERILIZATION
.
Dart Test cycle is preprogrammed, and cycle parameters are fixed and cannot
be changed by the customer.
Chamber must be at operating temperature when performing a Dart Test cycle.
The Dart Warmup cycle should be completed prior to performing the Dart Test.
Refer to "Dart Warmup Cycle Operation" included in this section.
1. Before running this cycle, refer to refer to
S
ECTION
6.1, B
EFORE
O
PERATING
S
TERILIZER
.
2. If necessary, prepare a test pack as described under "Constructing a Test
Pack" in
S
ECTION
3, T
ECHNIQUES
OF
S
TERILIZATION
.
3. Open chamber door and load chamber with Dart pack or prepared test pack.
NOTE: If sterilizer is equipped with double doors, the interlock type for Dart Test
cycle is factory set and fixed at Type #1. Refer to S
ECTION
8 for description of
interlock types.
4. Close chamber door.
5. Press
CYCLE SELECT
on screen #1.
6. Press
TEST CYCLES
button.
1.
PREVAC
4.
PREVAC
2.
GRAVITY
5.
GRAVITY
3.
LIQUID
6.
LIQUID
2
CYCLE SELECT
MAIN
MENU
TEST
CYCLES
Pressing MAIN MENU returns display to
screen #1.
Pressing OPTIONS advances display to
the first Out of Cycle Options menu
(screen #13).
OPTIONS
1
S T E R I S S C I E N T I F I C
STERILIZER PREPARED FOR:
STERIS SCIENTIFIC
CUSTOMER
** Not For Patient Use **
MAIN MENU
CYCLE
SELECT
OPEN
DOOR
CLOSE
DOOR