12-21
Troubleshooting
Operator Manual
129373-496
Door Switch Failure
(Continued)
Alarm
Description
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
238
DOOR SWITCH FAILURE
→
→
→
→
→
DOOR SWITCH OPEN WHILE SEAL SWITCH
CLOSED
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
→
→
→
→
CHECK LS1(LS2) CONNECTIONS
→
→
→
→
→
READJUST LS1(LS2)
→
→
→
→
→
REPAIR LS1(LS2)
2.
SEAL PRESSURE SWITCH MALFUNCTION
→
→
→
→
→
READJUST PS1(PS2)
→
→
→
→
→
REPAIR PS1(PS2)
PAPER
CLEAR
EXIT
FEED
ALARM
Screen with Operator Instructions
Occurs if liquid cycle
fast exhaust rate is too
fast.
Screen with Service Instructions
12.4.9 Exhaust Rate Too
Fast
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
241
EXHAUST RATE TOO FAST
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING
TO OPTIMAL COOLING RATE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
242
EXHAUST RATE TOO FAST
→
→
→
→
→
CHAMBER EXHAUSTED FASTER
THAN THE EXPECTED RATE.
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S40
2.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S03
3.
CHAMBER STEAM TRAP MALFUNCTION
→
→
→
→
→
REPAIR
4.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
CLEAR
EXIT
FEED
ALARM