Operating Instructions
Sterilizer Operation
6.4 Gravity Cycle
Operation
The Gravity cycle is designed for sterilizing heat- and moisture-stable goods
at 100° to 137°C (212° to 279°F).
1. Before running this cycle, refer to
S
ECTIONS
6.1, B
EFORE
O
PERATING
S
TERILIZER
and
6.2, L
OADING
/U
NLOADING
S
TERILIZER
.
2. Press
CYCLE SELECT
on screen #1.
WARNING – EXPLOSION HAZ-
ARD: This sterilizer is not de-
signed to process flammable
compounds.
3. Press
GRAVITY
button.
2
CYCLE SELECT
1.
PREVAC
4.
PREVAC
2.
GRAVITY
5.
GRAVITY
3.
LIQUID
6.
LIQUID
MAIN
MENU
TEST
CYCLES
Pressing TEST CYCLES advances
display to screen #3.
Pressing MAIN MENU returns display to
screen #1.
4. Verify cycle parameters listed for the selected Gravity cycle are acceptable.
If listed cycle parameters are not acceptable, press
PREVIOUS
button and
refer to
S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
, for instructions on changing
cycle parameters.
Pressing PREVIOUS returns display to
screen #2.
Pressing OPTIONS advances display to
the first Out-of-cycle Options menu
(screen #13).
6
CYCLE PARAMETERS
PREVIOUS
START
CYCLE
2. GRAVITY (GRAVITY)
PURGE TIME ......................................... 1:00
STERILIZE:
TEMP .................................................. 132.0 C
TIME ................................................... 0:15:00
VACUUM DRY ....................................... 10.0 inHg
DRY TIME .............................................. 0:30:00
Cycle Name
(Cycle Type)
OPTIONS