Troubleshooting
Operator Manual
129373-496
Pressure In Chamber
(Continued)
Screen with Operator Instructions
Screen with Service Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
WASTE:
000.0 C
STERILIZER WILL:
•
PREVENT NEW CYCLE FROM BEING
STARTED UNTIL ALARM IS CLEARED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
ALARM! WASTE
215
TEMPERATURE PROBE FAILURE
Alarm
Description
Screen with Service Instructions
12.3.4 Waste
Temperature
Probe Failure
Occurs if waste line
temperature reading
is outside the normal
range.
STATUS..
SERVICE INFORMATION:
222
PRESSURE IN CHAMBER
→
→
→
→
→
2 psig PRESSURE SENSED IN CHAMBER
WHEN NOT IN CYCLE
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S02
2.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3.
TRANSDUCER, CP, MALFUNCTION
→
→
→
→
→
REPAIR
→
→
→
→
→
RECALIBRATE
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
STATUS..
SERVICE INFORMATION:
216
WASTE TEMP PROBE FAILURE
→
→
→
→
→
RTD PROBE, RTD2, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTIONS IN PROBE WIRING
→
→
→
→
→
REPAIR
2.
PROBE FAILED
→
→
→
→
→
REPLACE
→
→
→
→
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE