11-3
Routine Maintenance
Operator Manual
129373-496
IMPORTANT:
The chamber drain strainer must be cleaned at least once a day,
preferably in the morning before running the first cycle.
1. Remove the drain strainer from the drain in the bottom of the chamber as shown
in Figure 11-1.
2. Remove any obvious debris from the strainer. If necessary, clear the screen
in the strainer using a brush, wire or similar tool.
3. Once it has been cleared of obvious debris, reverse flush the strainer under
running water.
4. Replace the strainer in the chamber drain.
11.1.1 Clean Chamber
Drain Strainer
W A R N I N G – B U R N
HAZARD: Allow sterilizer to
cool to room temperature
before performing any
cleaning or maintenance
procedures.
IMPORTANT:
The entire chamber should be wiped down and rinsed following
any spills or other soiling.
1. If applicable, the shelf assembly must be removed before cleaning the
chamber.
•
Single Door
a. Remove shelves from rack.
b. Use a
1
/
8
" hex wrench to loosen (but not remove) the set screws at the front
of the rack assembly.
c. Remove the rack assembly from the chamber.
•
Double Door
a. Remove shelves from rack.
b. Use a
1
/
8
" hex wrench to loosen (but not remove) the set screws at each
end of the rack assembly.
c. Remove the rack assembly from the chamber.
IMPORTANT:
Chamber must be at room temperature, sterilizer off all
night, before washing. Open the chamber emergency manual exhaust
valve. See Page 4-2 for location.
2. Wash the inside of the chamber and shelf assembly (plus any other loading
equipment) with a mild detergent solution such as STERIS Liqui-Jet
®
or
current STERIS equivalent. (
Contact your local STERIS representative.)
Figure 11-1. Remove Chamber Drain Strainer
11.1.2 Clean Chamber
WARNING - FALL HAZARD:
To prevent falls, keep
floors dry by immediately
wiping up any spilled
liquids or condensation in
sterilizer loading and
unloading areas.
WARNING - BURN HAZ-
ARD: Allow sterilizer to
cool to room tempera-
ture before performing
any cleaning or mainte-
nance procedures.
Chamber
Drain Strainer