Operating Instructions
Programming Operating Parameters
•
Pressing
EVACUATE
advances display to screen #43. Enter evacuate
phase value using the numeric keypad. Value appears on display as it is
being entered.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#40.
43
CHANGE VALUES - TOO LONG IN EVACUATE
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN EVACUATE
30:00
(Minutes:Seconds)
The too long in
evacuate is the
allowable amount of
time that the chamber
has to reach the
set vacuum level
before an alarm occurs.
4. Press
NEXT
button on screen #40. Screen #53 is the second of two menu
screens which list the phase and its currently set value. To change any of
the listed values, press the corresponding button.
NOTE: A definition of each value (parameter) is given on the corresponding
display screen.
53
CHANGE TOO LONG IN PHASE PARAMETERS
TOO LONG IN:
AIR BREAK = 30:00
DOOR SEAL = 30:00
DOOR UNSEAL = 30:00
JACKET CHARGE = 30:00
RETURN
PREVIOUS
MAIN
MENU
Pressing PREVIOUS returns display to
screen #40.
Pressing RETURN saves all changed
parameters and returns display to screen
#14.
Pressing MAIN MENU returns display to
screen #1.
•
Pressing
AIR BREAK
advances display to screen #44. Enter air break
phase value using the numeric keypad. Value appears on display as it is
being entered.
44
CHANGE VALUES - TOO LONG IN AIR BREAK
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN AIR
BREAK
30:00
(Minutes:Seconds)
The too long in
air break is the
allowable amount of
time that the chamber
has to reach 2 inHg
before an alarm occurs.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#53.