2-3
Installation Verification
Operator Manual
129373-496
2.2.3 Utility
Requirements
•
36" Sterilizer:
3800 lbs (1720 kg)
•
48" Sterilizer:
4200 lbs (1901 kg)
•
60" Sterilizer:
4700 lbs (2127 kg)
2.2.1 Overall Size
•
36" Sterilizer:
Hinged Door — 44" W x 51
1
/
2
" L x 75
1
/
4
" H
(1118mm W x 1308mm L x 1911 mm H)
Sliding Door — 70" W x 57" L x 75
1
/
4
" H
(1778 mm W x 1448 mm L x 1911 mm H)
•
48" Sterilizer:
Hinged Door — 44" W x 63
1
/
2
" L x 75
1
/
4
" H
(1118 mm W x 1600 mm L x 1911 mm H)
Sliding Door — 70" W x 69" L x 75
1
/
4
" H
(1778 mm W x 1753 mm L x 1911 mm H)
•
60" Sterilizer:
Hinged Door — 44" W x 75
1
/
2
" L x 75
1
/
4
" H
(1118 mm W x 1918 mm L x 1911 mm H)
Sliding Door — 70" W x 81" L x 75
1
/
4
" H
(1778 mm W x 2057 mm L x 1911 mm H)
2.2 Technical
Specifications
2.2.2 Weight
•
Electric (U.S.)
Controls
120 V, 1-phase, 2 A, 60 hz
Vacuum Pump
208/240 V, 3-phase, 6 A/phase, 60hz or
480 V, 3-phase, 3 A/phase, 60hz
•••••
Electric (Outside United Kingdom)
Controls
230 V, 1-phase, 1.5 A, 50 Hz
Vacuum Pump
400 V, 3-phase, 4 A/phase, 50 Hz
•
Electric (United Kingdom)
Controls
230 V, 1-phase, 1.5 A, 50 Hz
Vacuum Pump
400 V, 3-phase, 6 A/phase, 50 Hz
•
Water:
Pressure:
Sterilizer
20 to 50 psig (1.4 to 3.5 bar)
Temperature:
Sterilizer
70°F (21°C) maximum
Consumption:
Sterilizer
15 gpm (57 lpm), peak
(More)