12-2
129373-496
Operator Manual
Troubleshooting
Alarm
Description
Screen with Operator Instructions
Occurs if chamber does
not reach the set
temperature within the
allotted time.
Screen with Service Instructions
* ALARM
PRESSURE IN CHAMBER
F 10:07:23A
61.7C
2.34P
Full Print Format Shown
Figure 12-2. Typical Alarm
Printout
12.1.2 Typical Alarm
Printout
12.2 In-Cycle Alarms
12.2.1 Too Long In
Charge
→
CHAMBER DID NOT REACH STERILIZE
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 psig
→
CHECK STEAM SUPPLY PIPING
2.
STEAM REGULATOR MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
206
TOO LONG IN CHARGE
SERVICE
ABORT
EXIT
MODE
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
205
TOO LONG IN CHARGE
When an alarm occurs the printer automatically generates a printout, typically
listing alarm name, time alarm occurred, current chamber status, and any
associated sensor temperature. See Figure 12-2.
The following alarm screens will appear only during cycle operation: