8-12
129373-496
Operating Instructions
Programming Cycle Values
•
Pressing
VACUUM DRY
advances display to screen #37. Enter vacuum
dry value using the numeric keypad. Allowable vacuum dry range is 0-
29.9 inHg. Value appears on the display as it is being entered.
NOTE: If an out-of-range cycle value is entered, a display screen will
appear indicating the incorrect value and the allowable range. Display
screen will automatically return to previous screen, allowing operator to
enter an in-range cycle value.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#22.
37
CHANGE VALUES - VACUUM DRY
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET VACUUM DRY
10.0 inHg
The vacuum dry
parameter is the
vacuum level when
the dry time will start.
•
Pressing
DRY TIME
advances display to screen #38. Enter dry time
using the numeric keypad. Time appears on the display as it is being
entered.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#22.
38
CHANGE VALUES - DRY TIME
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET DRY TIME
00:05:00
(Hours:Minutes:Seconds)
The dry time is
the minimum time
spent in the dry
phase.