![Amsco Century Скачать руководство пользователя страница 146](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190146.webp)
11-1
Routine Maintenance
Operator Manual
129373-496
ROUTINE MAINTENANCE
Maintenance procedures described in Sections 11 and 13 must be performed
regularly at the indicated intervals, using the maintenance schedule in
Table 11-1 as a guide. Local conditions (water quality, usage, etc.) may
require more frequent maintenance than indicated. Refer to
Maintenance
Manual, P764330-117 for replacement parts list.
Customer should maintain a record of all maintenance procedures performed
on the sterilizer.
If a problem occurs, refer to
S
ECTION
12, T
ROUBLESHOOTING
.
NOTE: Never permit unqualified persons to service the sterilizer.
11.1 Preventive
Maintenance
Schedule
Table 11-1. Preventive Maintenance Schedule for Century Sterilizer
Service Required
Minimum Frequency
1.0
PREPARATION FOR PREVENTIVE MAINTENANCE
1.1
Discuss equipment with operators and check printouts.
6x per year
1.2
Follow appropriate safety procedures; prepare unit for preventive maintenance.
6x per year
2.0
DOOR ASSEMBLY (EACH DOOR ON A DOUBLE-DOOR UNIT)
2.1
Verify proper door and door proximity switch operation. Adjust switch(s)
if needed.
6x per year
2.2
Check condition of door gasket for wear and tear. Replace as needed.
6x per year
2.3
Verify proper tension on power door cable (sliding door units, only).
6x per year
3.0
VALVES
3.1
Verify each hand valve operates smoothly, check valve packing for leaks, rebuild
or replace as needed.
•
Steam supply valve.
6x per year
•
Water supply valve.
6x per year
3.2
Inspect check valves. Repair/replace as needed.
1x per year
3.3
Rebuild steam supply manifold.
1x per year
3.4
Rebuild all solenoid valves.
1x per year
3.5
Rebuild all check valves.
1x per year
3.6
Rebuild steam control valve (PRV).
1x per year
3.7
Verify proper setting/flow rate of flow control valves. Replace if needed.
2x per year
3.8
Verify that safety valve is not leaking.
6x per year
3.9
Replace safety valve.
1x per year
WARNING – SHOCK AND
BURN HAZARD: Regularly
s c h e d u l e d p r e v e n t i v e
maintenance is required for
safe and reliable operation
of this equipment. Contact
y o u r S T E R I S S e r v i c e
Representative to schedule
preventive maintenance.
11