4-4
129373-496
Operating Instructions
Component Identification
Ink-on-paper printer records all cycle data on 2-1/4" (57 mm) wide paper. The
generated printout is visible when the printer access door is closed (see
Figure 4-4).
All printer functions are controlled using the touch screen. For details on each
of the printer functions, refer to
S
ECTION
5, C
ONTROL
I
NTERFACE
.
The following is an example of a typical in-cycle printout in the
full print
format
(see Figure 4-5).
NOTE: Extended print format is available; refer to "Print Format" in S
ECTION
8.
•
Operating Mode
When sterilizer is placed in the Operating mode, the generated printout lists
the sterilizer type and manufacturer.
•
Cycle Start
When a cycle is started, the generated printout lists name of cycle started,
time and date the cycle was started, the current cycle count (number of
cycles run since original start up of unit), the operator's name, the sterilizer
ID number, the default cycle number and type, and the programmed
parameters for the cycle started.
NOTE: Cycle count value may be changed in service mode.
•
In-Cycle Performance
During a cycle, the generated printout lists the current time, chamber
pressure and chamber temperature at each transition point.
•
End-of-Cycle Performance Summary
At the end of a cycle, the generated printout lists number of cycles run that
day, the maximum and minimum chamber temperatures reached during
the sterilize phase, processing times for key phases, and the total cycle
time.
•
Alarm Condition
When an alarm condition occurs, the generated printout (see Figure 4-6)
lists the type of alarm and time, chamber temperature, and chamber
pressure when it occurred.
NOTE: Refer to S
ECTION
12, T
ROUBLESHOOTING
, for listing of possible alarm
conditions.
4.3.2 Printer
Operating
Mode
In-Cycle
Performance
Cycle
Start
Full Print Format Shown
Figure 4-5. Sample In-cycle
Printout
End-of-Cycle
Performance
Summary
Full Print Format Shown
Figure 4-6. Sample Alarm Printout
* ALARM
TOO LONG IN AIR BREAK
F 8:47:35A 70.1 25.0V