6-9
Sterilizer Operation
Operating Instructions
129373-496
• Charge
– After the last prevacuum pulse, chamber charges with steam to
the programmed sterilize temperature.
9
PREVAC CYCLE
PHASE:
CHARGE
CHARGING TO:
132.0 C
CHAMBER:
55.0 C
15.0 inHg
PROJECTED CYCLE COMPLETION TIME:
12:10
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Sterilize
–
Sterilize phase begins when chamber temperature is equal to
or greater than the programmed sterilize temperature. Chamber tempera-
ture is printed every 2 minutes (or printed after each programmed print
interval). The steam-to-jacket valve is regulated to maintain the chamber
at the sterilize control temperature (control temperature = sterilize tem-
pe overdrive temperature).
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
PREVAC CYCLE
PHASE:
STERILIZE
TIME LEFT IN PHASE:
4:00
CHAMBER:
132.0 C
29.0 psig
PROJECTED CYCLE COMPLETION TIME:
11:30
m:s
STATUS
ABORT
• Fast Exhaust
– Chamber is exhausted until chamber pressure reaches 4
psig (0.28 Pbar).
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
PREVAC CYCLE
PHASE:
FAST EXHAUST
FAST EXHAUSTING TO:
4.0 psig
CHAMBER:
133.5 C
30.0 psig
PROJECTED CYCLE COMPLETION TIME:
07:20
m:s
STATUS
ABORT