Sterilizer Operation
The Liquid cycle is designed for sterilizing liquids and media in vented
borosilicate glass or metal containers from 100° to 123°C (212° to 254°F).
1. Before running this cycle, refer to
S
ECTIONS
6.1, B
EFORE
O
PERATING
S
TERILIZER
and
6.2, L
OADING
/U
NLOADING
S
TERILIZER
.
2. Press
CYCLE SELECT
on screen #1.
6.5 Liquid Cycle
Operation
WARNING – EXPLOSION HAZ-
ARD: This sterilizer is not de-
signed to process flammable
compounds.
WARNING: It is inappropriate
for a healthcare facility to ster-
ilize liquids for direct patient
contact.
WARNING – BURN HAZARD:
When sterilizing liquids, you
must observe the following pro-
cedures:
• Use Liquid cycle only; no
other cycle is safe for pro-
cessing liquids.
• Use only vented closures.
• Use only Type I borosilicate
glass bottles.
• Do not allow hot bottles to
be jolted; this can cause hot-
bottle explosions.
3. Press
LIQUID
button.
2
CYCLE SELECT
1.
PREVAC
4.
PREVAC
2.
GRAVITY
5.
GRAVITY
3.
LIQUID
6.
LIQUID
MAIN
MENU
TEST
CYCLES
4. Verify cycle parameters listed for the selected Liquid cycle are acceptable.
If listed cycle parameters are not acceptable, press
PREVIOUS
button and
refer to
S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
, for instructions on changing
cycle parameters.
Pressing TEST CYCLES advances
display to screen #3.
Pressing MAIN MENU returns display to
screen #1.
Pressing PREVIOUS returns display to
screen #2.
5
CYCLE PARAMETERS
PREVIOUS
START
CYCLE
3. LIQUID (LIQUID)
PURGE TIME ......................................... 1:00
STERILIZE:
TEMP .................................................. 121.0 C
TIME ................................................... 0:45:00
Cycle Name
(Cycle Type)
OPTIONS
1
S T E R I S S C I E N T I F I C
STERILIZER PREPARED FOR:
STERIS SCIENTIFIC
CUSTOMER
** Not For Patient Use **
MAIN MENU
CYCLE
SELECT
OPEN
DOOR
CLOSE
DOOR