11-2
129373-496
Operator Manual
Routine Maintenance
Service Required
Minimum Frequency
4.0
MISCELLANEOUS PIPING COMPONENTS
4.1
Inspect steam strainer for debris, clean as needed.
2x per year
4.2
Inspect water strainer for debris, clean as needed.
2x per year
4.3
Inspect jacket strainer for debris, clean as needed.
2x per year
4.4
Inspect chamber drain strainer for debris, clean as needed.
6x per year
4.5
Replace air filter cartridge.
1x per year
4.6
Chamber and jacket gauge(s) - verify proper operation. Replace if needed.
6x per year
4.7
Rebuild chamber and jacket traps.
1x per year
4.8
Verify that there are no leaks.
6x per year
4.9
Verify that door lock piston operates correctly.
6x per year
5.0
CONTROL
5.1
Verify that printer and paper take-up operate properly.
Check printout for darkness, missing dots, etc.
6x per year
5.2
Verify that all touch panels function properly (O.E. & N.O.E).
6x per year
5.3
Verify that the date and time are correct. If not, correct.
6x per year
5.4
Verify operation of the battery-backed RAM, replace as needed.
6x per year
5.5
Verify that the buzzer is working.
6x per year
5.6
Verify that the water level sensor operates properly.
6x per year
5.7
Verify that cooling fan operates properly.
6x per year
5.8
Replace fan filter.
1x per year
5.9
Check all service-settable values in Service Test Mode for factory
recommended settings. Verify functional operation of each valve using
the Service Test Mode.
1x per year
5.10 Verify temperature displays/printouts with potentiometer.
6x per year
5.9
Verify temperature and pressure settings as described in Maintenance Manual.
1x per year
6.0
SAFETY TESTING
6.1
Inspect ground bond.
1x per year
6.2
Inspect steam connection to sterilizer.
1x per year
6.3
Inspect water connection to sterilizer.
1x per year
6.4
Inspect drain connection to sterilizer.
1x per year
6.5
Test ground continuity to sterilizer per EN 61010-1.
1x per year
7.0
FINAL CHECKOUT AND TEST
7.1
Clean dirt and lint from components. Check all wiring, terminals, and
socket connections for damage or fraying.
6x per year
7.2
Verify that unit has proper labels (caution, warning).
6x per year
7.3
Run machine through each cycle to verify proper operation. Check all
displays and printouts. Note on tape, “STERIS TEST."
6x per year
7.4
Verify that the shelves slide easily in and out of the chamber, if equipped.
6x per year
7.5
Reinstall any panel or cover removed. Check area to ensure removal of all
materials used during inspection.
IMPORTANT:
The entire chamber should be wiped down and rinsed following any spills or other soiling.
Table 11-1. Preventive Maintenance Schedule for Century Sterilizer (continued)