12-1
Troubleshooting
Operator Manual
129373-496
This section pictorially lists and describes all the possible alarm condi-
tions which may occur when operating the Amsco
®
Century
TM
Medium
Steam Sterilizer 26" x 37.5".
If a problem occurs that is not described in this section, please call your
STERIS Service representative. A trained service technician will promptly
place your sterilizer in proper working condition.
NOTE: Never permit unqualified persons to service the sterilizer.
12.1 General
When an alarm condition occurs, the alarm tone sounds and the touch
screen automatically displays the corresponding alarm screen. Typically,
each alarm screen indicates the alarm name, current chamber status,
current sterilizer activity, and operator instructions (see Figure 12-1).
Touch-screen pads, located along bottom of alarm screen, are used to perform
the following functions:
•
Pressing
SILENCE ALARM
turns off the alarm tone.
•
Pressing
STATUS PRINT
generates a printout of the current temperature
and pressure in the sterilizer chamber at the time the touch pad was
pressed.
•
Pressing
PAPER FEED
advances the printer paper up by one line.
•
Pressing
SERVICE HELP
advances display to the corresponding service
information screen. This screen provides the qualified service technician
with possible causes and advanced corrective actions for that alarm
condition.
IMPORTANT:
In the event of an alarm condition, the operator should always
follow the instructions indicated on the alarm screen.
TROUBLESHOOTING
12.1.1 Typical Alarm
Screen
WARNING – PERSONAL
INJURY AND/OR EQUIP-
MENT DAMAGE HAZARD:
Repairs and adjustments
to this equipment must
be made only by fully
qualified service person-
nel. Maintenance per-
formed by inexperienced,
unqualified persons or
installation of unautho-
rized parts could cause
personal injury or result
in costly equipment dam-
age.
WARNING–SHOCK HAZ-
ARD: Disconnect all utilities
to sterilizer before servicing.
Do not service the sterilizer
unless all utilities have been
properly locked out. Always
follow OSHA Lockout-
Tagout and electrical safety-
related work practice stan-
dards. (See CFR 1910.147
and .331 through .335.).
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
205
TOO LONG IN CHARGE
Figure 12-1. Typical Alarm Screen
Screen Reference Number
Sterilizer Response to
Condition
Current Chamber
Status
Alarm Name
Operator Instructions
Touch-Screen Pads
12
WARNING–BURN HAZARD:
Allow sterilizer to cool to
room temperature before per-
forming any cleaning or
maintenance procedures.