Programming Cycle Values
5. Press
NEXT
button on screen #20. Screen #21 is the second of three menu
screens which list the programmable values for the Prevacuum cycle type.
To change any of the listed cycle values, press the corresponding button.
NOTE: A definition of each cycle value is given on the corresponding display
screen.
21
CHANGE VALUES - 1. PREVAC (PREVAC)
STER. TIME = 4:00
STER. TEMP = 132.0 C
OVERDRIVE = 1.5 C
UNDER TEMP = 1.0 C
OVER TEMP = 6.0 C
NEXT
PREVIOUS
MAIN
MENU
•
Pressing
STER. TIME
advances display to screen #32. Enter sterilize
time using the numeric keypad. Time appears on display as it is being
entered.
Pressing PREVIOUS returns display to
screen #20.
Pressing NEXT advances display to
screen #22.
Pressing MAIN MENU returns display to
screen #1.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#21.
32
CHANGE VALUES - STER. TIME
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET STER. TIME
00:04:00
(Hours:Minutes:Seconds)
The ster. time is
the minimum time
spent during the
sterilize phase.
The actual sterilize
time may become longer because of
undertemp or overtemp alarms.
Cycle Name
(Cycle Type)