![Amsco Century Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190118.webp)
10-7
Out of Cycle Options
Operating Instructions
129373-496
4. The second values screen appears listing more cycle values programmed
for the selected cycle.
If applicable, press
NEXT
to view remaining programmed values for the
selected cycle.
NOTE: Depending on cycle selected, there are either two or three values
screens showing the programmed cycle values.
97
DISPLAY VALUES
1. PREVAC (PREVAC)
STER. TIME .........................................
4:00
STER. TEMP ........................................
132.0 C
OVERDRIVE ........................................
1.5 C
UNDER TEMP ......................................
1.0 C
UNDER TEMP ......................................
RESTART
OVER TEMP ........................................
6.0 C
PRINT INT ............................................
2 min
Pressing PREVIOUS returns display to
first values screen shown.
Pressing MAIN MENU returns display to
screen #1.
PREVIOUS
MAIN
MENU
NEXT
5. The last values screen appears listing the remaining cycle values pro-
grammed for the selected cycle.
After viewing the cycle values, press
CYCLE MENU
on the last values
screen. Display returns to screen #93.
98
DISPLAY VALUES
1. PREVAC (PREVAC)
VACUUM DRY .....................................
10.0 inHg
DRY TIME .............................................
30:00
INTERLOCK .........................................
2
PREVIOUS
MAIN
MENU
CYCLE
MENU
Pressing PREVIOUS returns display to
second values screen shown.
Pressing MAIN MENU returns display to
screen #1.
Cycle Name
(Cycle Type)
Double Door
Units Only
Cycle Name
(Cycle Type)