Operating Instructions
Sterilizer Operation
4. Verify cycle parameters listed for the selected Prevacuum cycle are
acceptable.
If listed cycle parameters are not acceptable, press
PREVIOUS
button and
refer to
S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
, for instructions on changing
cycle parameters.
5. To begin Prevacuum cycle operation, press
START CYCLE
.
NOTE: If START CYCLE is pressed while the chamber door is open, a
display screen will appear directing the operator to close door before
continuing cycle operation. Operator must reselect the cycle after door is
closed.
NOTE: If a cycle is started when the sterilizer has not been fully calibrated,
a display screen will appear indicating that the control is not calibrated.
Sterilizer must be calibrated by a qualified service technician before process-
ing loads.
6. Sterilizer automatically progresses through the following cycle phases.
NOTE: If an alarm occurs during cycle operation, refer to
S
ECTION
12,
T
ROUBLESHOOTING
,
for instructions on correcting the alarm condition.
NOTE: If power is lost during cycle operation, cycle either continues in same
phase or aborts if seal pressure is below 5 psig (0.34 bar) once power is
restored.
• Jacket Charge
– Jacket charges with steam to 1° less than the pro-
grammed sterilize temperature.
4
CYCLE PARAMETERS
PREVIOUS
START
CYCLE
1. PREVAC (PREVAC)
PURGE TIME ......................................... 1:00
PULSES ................................................. 4
MAX: 26.0 psig MIN: 10.0 inHg
STERILIZE:
TEMP .................................................. 132.0 C
TIME ................................................... 0:04:00
VACUUM DRY ....................................... 10.0 inHg
DRY TIME .............................................. 0:05:00
Cycle Name
(Cycle Type)
Pressing PREVIOUS returns display to
screen #2.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
PREVAC CYCLE
PHASE:
JACKET CHARGE
JACKET CHARGING TO:
120 C
CHAMBER:
32.0 C
00.0 psig
PROJECTED CYCLE COMPLETION TIME:
33:10
m:s
ABORT
STATUS