![Amsco Century Скачать руководство пользователя страница 171](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190171.webp)
12-16
129373-496
Operator Manual
Troubleshooting
The following alarm screens will appear anytime the sterilizer is energized.
The sensors are continually monitored whenever the sterilizer is in or out of
cycle.
12.4.1 Water In
Chamber
Occurs if excess water
is sensed in the
chamber.
Screen with Service Instructions
Screen with Operator Instructions
Occurs if jacket does
not reach set tempera-
ture within allotted time.
12.4.2 Too Long In
Jacket Charge
12.4 Sensor Alarms
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
231
TOO LONG IN JACKET CHARGE
CHAMBER:
000.0 C
0.0 psig
JACKET:
000.0 C
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
JACKET TEMP. IS REACHED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
→
→
→
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
WARNING - BURN HAZARD!
CHAMBER MAY BE FILLED WITH STEAM
CONDENSATE
OPERATOR INSTRUCTIONS:
1. DO NOT OPEN DOOR
2. SILENCE ALARM
3. CALL SERVICE IMMEDIATELY
STATUS..
ALARM!
209
WATER IN CHAMBER
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
210
WATER IN CHAMBER
→
→
→
→
→
EXCESS WATER SENSED IN CHAMBER,
WARNING! BURN HAZARD
CAUSES AND CORRECTION:
1.
JACKET STRAINER PLUGGED
→
→
→
→
→
CLEAN
2.
JACKET TRAP FAILED CLOSED
→
→
→
→
→
REPAIR
3.
CHAMBER TRAP FAILED CLOSED
→
→
→
→
→
REPAIR
4.
WATER ENTERED CHAMBER THROUGH STEAM
PIPING
→
→
→
→
→
CHECK BOILER OR STEAM GENERATOR
→
→
→
→
→
REPAIR PIPING
5.
WATER FLOAT SENSOR MALFUNCTION
→
→
→
→
→
REPAIR
PAPER
SERVICE
EXIT
FEED
MODE