8-9
Programming Cycle Values
Operating Instructions
129373-496
•
Pressing
STER. TEMP
advances display to screen #33. Enter sterilize
temperature using the numeric keypad. Allowable sterilize temperature
ranges are 121°-137°C (250°-279°F) for Prevac cycles; 100°-137°C
(212°-279°F) for Gravity cycles; and 100°-123°C (212°-254°F) for Liquid
cycle. Temperature appears on display as it is being entered.
NOTE: If an out-of-range cycle value is entered, a display screen will
appear indicating the incorrect value and the allowable range. Display
screen will automatically return to previous screen, allowing operator to
enter an in-range cycle value.
•
Pressing
OVERDRIVE
advances display to screen #34. Enter overdrive
temperature using the numeric keypad. Temperature appears on display
as it is being entered.
34
CHANGE VALUES - OVERDRIVE
SET OVERDRIVE
0.8 C
The overdrive plus
the sterilize temp.
is the control temp.
during the sterilize
phase.
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
33
CHANGE VALUES - STER. TEMP
SET STER. TEMP
132.0 C
The sterilize temp.
is the temp. when
the sterilize phase will start.
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#21.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#21.