6-22
129373-496
Operating Instructions
Sterilizer Operation
• Slow Evacuate
– A vacuum is slowly drawn in the chamber to 5.0 inHg.
This phase ensures that the chamber is cooled to 95°C (203°F).
9
LIQUID CYCLE
PHASE:
SLOW EVACUATE
EVACUATING TO:
5.0 inHg
CHAMBER:
95.5 C
0.0
psig
PROJECTED CYCLE COMPLETION TIME:
11:20
m:s
STATUS
ABORT
• Vapor Removal
– Filtered air enters the chamber to relieve the vacuum
within the chamber. Steam is exhausted from the door seal. Chamber
vapor vents, through slight opening between seal and door, into the
sterilizer cabinet for 6 minutes.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
LIQUID CYCLE
PHASE:
VAPOR REMOVAL
TIME LEFT IN PHASE:
6:00
CHAMBER:
90.0 C
0.2
inHg
PROJECTED CYCLE COMPLETION TIME:
05:40
m:s
STATUS
ABORT
• Complete
– The complete tone sounds and the cycle summary and end-
of-cycle messages are printed.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
7
LIQUID CYCLE
PHASE:
COMPLETE
OPEN DOOR(S) TO UNLOAD
CHAMBER:
32.0 C
00.0 psig
W A R N I N G
H O T L I Q U I D S !
STATUS
ABORT