12-19
Troubleshooting
Operator Manual
129373-496
Screen with Operator Instructions
Chamber Pressure
Transducer Failure
(Continued)
Occurs if chamber
temperature reading is
outside the normal
range.
12.4.6 Chamber
Temperature
Probe Failure
Screen with Service Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! CHAMBER
211
TEMPERATURE PROBE FAILURE
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Alarm
Description
Screen with Service Instructions
STATUS..
SERVICE INFO: CHAMBER
218
PRESSURE TRANSDUCER FAILURE
→
→
→
→
→
TRANSDUCER, CP, OUTPUT VOLTAGE IS
OUTSIDE NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN TRANSDUCER
WIRING
→
→
→
→
→
REPAIR
2.
TRANSDUCER FAILED
→
→
→
→
→
REPAIR
→
→
→
→
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
STATUS..
SERVICE INFORMATION:
2 1 2
CHAMBER TEMP PROBE FAILURE
→
→
→
→
→
RTD PROBE, RTD1, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
→
→
→
→
REPAIR
2.
PROBE FAILED
→
→
→
→
→
REPLACE
→
→
→
→
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE