8-4
129373-496
Operating Instructions
Programming Cycle Values
The Change Values procedure is used to change preset cycle values
(parameters) for a particular cycle. The following procedure is an example of
the Change Values procedure for a Prevacuum cycle. The cycle phase values
depicted on the touch screen examples are the settings of the default
Prevacuum cycle.
The procedures for changing the Gravity and Liquid cycle values are similar,
with the exception that some programmable Gravity and Liquid cycle values
are different.
IMPORTANT NOTE:
If preset cycle values are changed, it is necessary for the
operator to validate the efficacy of the changed cycle.
1. Access Cycle Menu - Change Cycle Values (screen #15) as described at
the beginning of this section.
2. Press
PREVAC
button.
8.2 Change Values
Procedure
Pressing RETURN returns display to
screen #14.
Pressing MAIN MENU returns display to
screen #1.
15
CYCLE MENU - CHANGE CYCLE VALUES
RETURN
MAIN
MENU
1.
PREVAC
4.
PREVAC
2.
GRAVITY
5.
GRAVITY
3.
LIQUID
6.
LIQUID
3. Screen #16 allows operator to choose the type of cycle that will be assigned
to the selected cycle button. The current cycle name and assigned cycle
type are shown at the bottom of the display.
For this example, press
PREVAC
to assign a Prevacuum cycle type.
16
CHANGE CYCLE TYPE
CURRENT TYPE: PREVAC
NAME: 1. PREVAC
CYCLE
MENU
Pressing CYCLE MENU returns display
to screen #15.
PREVAC
GRAVITY
LIQUID