6-15
Sterilizer Operation
Operating Instructions
129373-496
• Charge
– Chamber charges with steam to the programmed sterilize
temperature.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
GRAVITY CYCLE
PHASE:
CHARGE
CHARGING TO:
132.0 C
CHAMBER:
32.0 C
00.0
psig
PROJECTED CYCLE COMPLETION TIME:
46:10
m:s
STATUS
ABORT
• Sterilize
– Sterilize phase begins when chamber temperature is equal to
or greater than the programmed sterilize temperature. Chamber tempera-
ture is printed every 2 minutes (or printed after each programmed print
interval). The steam-to-jacket valve is regulated to maintain the chamber
at the sterilize control temperature (control temperature = sterilize tem-
pe overdrive temperature).
9
GRAVITY CYCLE
PHASE:
STERILIZE
TIME LEFT IN PHASE:
15:00
CHAMBER:
132.0 C
29.0
psig
PROJECTED CYCLE COMPLETION TIME:
44:00
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Fast Exhaust
– Chamber is exhausted until chamber pressure reaches
4 psig (0.28 Pbar).
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
GRAVITY CYCLE
PHASE:
FAST EXHAUST
FAST EXHAUSTING TO:
4.0 psig
CHAMBER:
133.5 C
30.0
psig
PROJECTED CYCLE COMPLETION TIME:
38:50
m:s
STATUS
ABORT