![Amsco Century Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190015.webp)
2-4
129373-496
Operator Manual
Installation Verification
2.2.4 Environmental
Conditions
•
Steam:
Pressure:
50 to 80 psig (3.5 to 5.2 bar)
Consumption:
36" Sterilizer:
190 lb/hr (86 kg/hr)
48"
Sterilizer:
255 lb/hr (116 kg/hr)
60"
Sterilizer:
335 lb/hr (152 kg/hr)
Temperature: 50° to 90°F (10° to 32°C)
Humidity: 10 percent to 90 percent noncondensing
Pollution Degree: 2
Installation Category (Overvoltage Category): II
A-Weighted Sound Power Level:
≤
80 dBA (maximum)