<4. YH8000 Installation>
4-7
IM 11Y01D01-01EN
The factory default setting is for the cables to come out from the right side.
INV
NORM
YH8000 inversion switch
YH8000 inversion switch
How the YH8000 is mounted
Inversion switch position
Cables coming out from the right side.
NORM
Cables coming out from the left side.
INV
4.3
Remote HMI Installation
This section explains how to mount the YH8000 using mounting brackets (pipe mount (option
code: /P), wall mount (option code: /W)).
The sun shield (option code: /S) for avoiding the effects of direct sunlight can also be mounted.
WARNING
The TDLS8000 does not have a power switch. Provide a switch on the power supply line to
separate the TDLS8000 from the main power supply. Use labels to indicate that this switch is for
cutting off the power supply to the TDLS8000 and to indicate ON and OFF.
l
Pipe mounting
The YH8000 can be mounted on a 50A (2B) pipe. Fasten the YH8000 bolts with a torque of about
5 to 6 N•m. The YH8000 can be mounted horizontally or vertically.
The pipe mount option includes short bolts that are used when a sun shield is used and long bolts
that are used when a sun shield is not used.
Washer
Vertical mounting
Horizontal mounting
Long bolt for
pipe mounting
Figure 4.5
Pipe mounting
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...