![Vigilantplant TDLS8000 User Manual Download Page 139](http://html1.mh-extra.com/html/vigilantplant/tdls8000/tdls8000_user-manual_1022489139.webp)
<6. Configuration>
6-24
IM 11Y01D01-01EN
l
Offline validation (*1)
Parameter
Initial value
Min. – Max.
Pressure mode for offline validation 1
Process parameter Select in the screen
Temperature mode for offline validation 1
Process parameter Select in the screen
OPL mode for offline validation 1
Process parameter Select in the screen
Pressure fixed value for offline validation 1
101.325[kPa]
0.1 – 10,000[kPa]
Temperature fixed value for offline validation 1
25[deg C]
-273 – 3,000[deg C]
OPL fixed value for offline validation 1
0.66[m]
0.01 – 100[m]
Gas1 concentration value for offline validation 1
200,000[ppm]
0 – 1E6[ppm](*2)
Gas2 concentration value for offline validation 1
200,000[ppm]
0 – 1E6[ppm](*2)
Gas type for offline validation 1
Gas 1
Select in the screen
Auto offline validation 1 time initiate
Disable
Select in the screen
Auto offline validation 1 time initiate cycle (day)
0 (=Disable)
0 – 999
Auto offline validation 1 time initiate cycle (hour)
0 (=Disable)
0 – 23
Auto offline validation 1 time initiate base clock
2010/01/01 00:00:00 2010/01/01 00:00:00 –
2068/12/31 23:59:59
Auto offline validation 1 gas purge time
600[s]
0 – 10,000[s]
Auto offline validation 1 process purge time
600[s]
0 – 10,000[s]
Auto valve control for manual offline validation 1
Disable
Select in the screen
Auto offline validation 1+2 time initiate
Disable
Select in the screen
Auto offline validation 1+2 time initiate cycle (day)
0 (=Disable)
0 – 999
Auto offline validation 1+2 time initiate cycle (hour)
0 (=Disable)
0 – 23
Auto offline validation 1+2 time initiate base clock
2010/01/01 00:00:00 2010/01/01 00:00:00 –
2068/12/31 23:59:59
*1:
The initial value for the parameters related to “offline validation 2” is the same as those for “offline validation 1.”
*2:
Zero is not allowed.
l
Online validation (*1)
Parameter
Initial value
Min. – Max.
Temperature mode for online validation 1
Active ambient
Select in the screen
Temperature ambient offset for online validation 1
-2.2
[deg C]
-100 – 100
[deg C]
Pressure value for online validation 1
101.325[kPa]
0.1 – 10,000[kPa]
Temperature fixed value for online validation 1
25
[deg C]
-273 – 3,000
[deg C]
OPL value for online validation 1
0.1306[m](*2)
0.01 – 10[m]
Gas1 concentration value for online validation 1
200,000[ppm]
-1E6 – 1E6[ppm]
Gas2 concentration value for online validation 1
200,000[ppm]
-1E6 – 1E6[ppm]
Gas type for online validation 1
Gas 1
Select in the screen
Auto online validation 1 time initiate
Disable
Select in the screen
Auto online validation 1 time initiate cycle (day)
0 (=Disable)
0 – 999
Auto online validation 1 time initiate cycle (hour)
0 (=Disable)
0 – 23
Auto online validation 1 time initiate base clock
2010/01/01 00:00:00 2010/01/01 00:00:00 –
2068/12/31 23:59:59
Auto online validation 1 validation gas purge time
600[s]
0 – 10,000[s]
Auto online validation 1 normal gas purge time
600[s]
0 – 10,000[s]
Auto valve control for manual online validation 1
Disable
Select in the screen
Reading mode for online validation 1
Validation
Select in the screen
Concentration output factor during online validation 1
1.0
-9.9 – 9.9
*1:
The initial value for the parameters related to “online validation 2” is the same as those for “online validation.”
*2:
The parameter may be initialized to a value different from the one shown because the initial value differs depending on the
measurement target gas (application) of the TDLS8000.
l
Concentration offset
Parameter
Initial value
Min. – Max.
Concentration offset for gas1
0[ppm]
-1E6 – 1E6[ppm]
Concentration offset for gas2
0[ppm]
-1E6 – 1E6[ppm]
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...