<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-3
IM 11Y01D01-01EN
3.1.2
Constructing Process Flanges
Please prepare your own process flanges.
n
Process Flange Reinforcing Plate
If the duct or the wall of flue that the process flanges will be attached to is thin or may bend, weld
large reinforcing plates around the area where the flanges will be attached. Figure 3.1 shows an
example of welding reinforcing plates. Provide a secure platform to install the TDLS8000 on at
your discretion.
Wall containing the measured gas
Side view
Front view
Approx. 10 mm thick steel plate (welded)
Reinforcing plate
(welded)
LU or SCU
Figure 3.1
Reinforcing plate for LU and SCU process flanges
CAUTION
Duct or flue with thin and easily bent walls must be reinforced to maintain laser beam alignment
at all times. Reinforce the process flanges to maintain alignment.
3rd Edition: Jun. 10, 2016-00