<9. Inspection and Maintenance>
9-13
IM 11Y01D01-01EN
(3) Discharging the check gas
If automatic valve control is disabled, manually control the valves to purge the validation
cell with the analyzer internal purge gas (nitrogen gas) that is used during process
measurement. Check that the concentration is stable, and then touch Execute. The
validation result will be displayed.
(4) Checking the online validation result
The online validation result is displayed, and online validation ends. If successful, “PASSED”
appears. Otherwise, “FAILED” appears.
Expected Reading:
Gas concentration (expected value) obtained by adding the
check gas
Actual Reading:
The actual value
Tap OK to return to the configuration menu.
NOTE
If the validation fails, the following warning will occur. For the corrective action, see “10.2 Warning
Alarm number
Alarm name
15
Validation Error
3rd Edition: Jun. 10, 2016-00