<5. Startup>
5-20
IM 11Y01D01-01EN
5.3.6
Setting Process Alarms
You can set threshold values of the high/low limit alarm (warning) for process measurement
values. The following table shows the types of warnings that you can specify. For each type of
warning, you can set the threshold values and select whether to enable the detection. For details
on warnings, see “10.2 Warning Display and Handling” described later.
Alarm number
Warning name
1
Transmission Low
2
Process Pressure Low
3
Process Pressure High
4
Process Temperature Low
5
Process Temperature High
6
Concentration Gas1 Low
7
Concentration Gas1 High
8
Concentration Gas2 Low (only with two-gas measurement)
9
Concentration Gas2 High (only with two-gas measurement)
l
Setup procedure using HART
(1) From the menu, select “Detailed setup>>Alarm>>Warning>>Warning group 1.”
The following screen will appear.
(2) “Warn select group 1” is used to enable or disable the warnings. By factor default, all
warnings are enabled.
Select “Warn select group 1” and press the right arrow button to display the following
screen. Here, you can enable or disable each warning.
ON means enabled. To switch between ON and OFF, tap OFF (center-left at the bottom of
the screen). When you are finished, tap ENTER.
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...