![Vigilantplant TDLS8000 User Manual Download Page 33](http://html1.mh-extra.com/html/vigilantplant/tdls8000/tdls8000_user-manual_1022489033.webp)
<2. Specifications>
2-8
IM 11Y01D01-01EN
l
General Specifications
Process connections: (See next table)
Heatresistance temperature: 200ºC max
Measured gas pressure:
Max. 1 MPa abs.
Wetted materials: Sapphire, 316 SS, Monel 400, Kalrez (O-ring)
Weight;
Process connection
Analyzer connection
Weight
316SS
Monel 400
ANSI Class 150-2-RF Flange
ANSI Class 150-2-RF Flange
5 kg/pc
6 kg/pc
ANSI Class 300-2-RF Flange
7 kg/pc
7 kg/pc
ANSI Class 150-3-RF Flange
8 kg/pc
9 kg/pc
ANSI Class 300-3-RF Flange
11 kg/pc
12 kg/pc
ANSI Class 150-4-RF Flange
12 kg/pc
14 kg/pc
DIN PN16-DN50 Flange
DIN PN16-DN50 Flange
7 kg/pc
7 kg/pc
DIN PN16-DN80 Flange
10 kg/pc
11 kg/pc
JIS 10K-50-FF Flange
7 kg/pc
7 kg/pc
JIS 10K-80-FF Flange
9 kg/pc
10 kg/pc
Note
:
When using TDLS8000 as CE marking compliance product, the upper limit of the measurement gas pressure is 50kPa in gauge
pressure.
2.2.3
YC8000 Flow Cell
Used for extracting sample streams at any location.
Note: Must use in conjunction with alignment flanges (“-FC”)
l
General Specifications
Gas temperature: 200ºC max
Gas pressure: Max. 1 MPa abs.
Wetted materials: Sapphire, 316 SS, Monel 400, Kalrez (O-ring)
Weight;
Material/Optical Path Length 1016 mm (40 inch) 1524 mm(60 inch)
Monel 400
15 kg
18 kg
316SS
14 kg
17 kg
Note: When using TDLS8000 as CE marking compliance product, the upper limit of the measurement gas pressure in YC8000 is
50kPa in gauge pressure.
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...