<11. Modbus>
11-3
IM 11Y01D01-01EN
The error code details are provided below.
Error code
Description
01
Function code error (nonexistent function)
02
Coil, input relay, or register address error (out of range)
03
Coil, input relay, or register data number error (out of range)
06
During instruction message execution, an error which is the slave device
cannot execute occurs.
Example: Writing not possible because maintenance is in progress
07
Command error
Example: Write-data is out of range.
11.2 Coil
Coil name
Address
Action performed when “1” is set
Automatic zero calibration execution
00001 Remotely execute an automatic zero calibration
Automatic span calibration execution
00002 Remotely execute an automatic span calibration
Automatic offline validation 1 execution
00004
Remotely execute automatic offline validation 1
Automatic offline validation 2 execution
00005
Remotely execute automatic offline validation 2
Automatic online validation 1 execution
00006 Remotely execute automatic online validation 1
Automatic online validation 2 execution
00007 Remotely execute automatic online validation 2
Time set instruction
00008 Set the hold register time value (40201 to 40206)
Automatic zero+span calibration
execution
00009
Remotely execute an automatic zero+span calibration
Automatic offline validation 1+2
execution
00010
Remotely execute automatic offline validation 1+2
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...