<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-17
IM 11Y01D01-01EN
3rd Edition: Jun. 10, 2016-00
CAUTION
Check that the TDLS8000 is turned off before setting SW1 or SW2. Otherwise, the TDLS8000
may be damaged.
n
Connecting a Pressure Meter and Thermometer
Connect the analog output terminals of the transmitters as follows. The terminal polarity is the
same for Passive AI and Active AI.
AO-1
AI-2
AO-2 AI-1
DI-1
+
+ -
+ -
-
FAULT
DI-2
DO
+
-
Internal
ground terminal
Temperature transmitter
Sensor control unit
Terminal B
Pressure transmitter
n
When Connecting an External Power Supply Such as a Distributor
If you need to connect an external power supply such as a distributor to a 2-wire system
transmitter, connect it as follows. Set the switch to Passive AI.
AO-1
AI-2
AO-2 AI-1
DI-1
FAULT
DI-2
DO
+
+
-
-
Internal
ground terminal
Sensor control unit
Terminal B
External power supply
24V DC
Pressure transmitter
l
Wiring procedure
• For the cable type to use, see “
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• When supplying power to the transmitters from the TDLS8000, take into account the drop in
the transmitter supply voltage due to wiring resistance and the like.
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...