<3. Installation, Wiring, Optical Axis Adjustment, and Piping>
3-20
IM 11Y01D01-01EN
3.2.6
Wiring Digital Inputs
The TDLS8000 executes specified functions when it receives contact signals. There are two
inputs. Apply voltage-free contact signals. The digital input terminal outputs 5 V DC.
For digital input settings, see “6.7 Digital Input Settings”.
AO-1
AI-2
AO-2 AI-1
DI-1
+ -
+ -
FAULT
DI-2
DO
SW1
SW2
Internal ground terminal
Sensor control unit
Terminal B
l
Wiring procedure
• For the cable type to use, see “
• Be sure to ground the cable shield on the TDLS8000 side and on the other side.
• The open and closed levels of the digital inputs are identified by the resistance seen from
the TDLS8000 side. Wire the digital inputs to meet the following conditions. Note that wiring
resistance is included.
If the DI terminals are shorted, about 2 mA of current will flow.
Closed
Open
Resistance
200Ω or less
100 kΩ or more
3.2.7
Wiring Solenoid Valve Control Outputs
These outputs control the solenoid valves that are used during calibration and the like. There are
two outputs. Each can supply 24 V DC 500 mA max.
LC
VO
MS-1 MS-2
SV-1
VO(HMI)
SV-2 POW
Internal ground terminal
Connect to the TDLS8000
shielded wire terminal.
Sensor control unit
Terminal C
Solenoid valve 1
Solenoid valve 2
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...