![Vigilantplant TDLS8000 User Manual Download Page 44](http://html1.mh-extra.com/html/vigilantplant/tdls8000/tdls8000_user-manual_1022489044.webp)
<2. Specifications>
2-19
IM 11Y01D01-01EN
n
YC8000 Flow Cell
TDLS8000 have to be assigned the dedicated Alignment flange (Optic Accessory: -FC).
When piping is Rc1/4, a conversion adopter will be attached on the Alignment flange.
Ø152.4
35
Ø152.4
Unit: mm
Sample inlet
(for 1/4” pipe)
Sample outlet
(for 1/4” pipe)
Gas temp. or
pressure measurement port
(for 1/4” pipe)
Pipe size: 1-1/2 Sch80
Flow cell aligenment
flange
(TDLS8000
Optics accessory
code “-FC”)
86 (vaeiable)
1053 (Optical path length: -40) or 1561 (Optical path length: -60)
n
Calibration Cell
Part number: K9772XA, K9772XB, K9772XC, K9772XD, K9772XE, K9772XE, K9772XF,
K9772XG, K9772XH, K9772XJ, K9772XL, K9772XM
Ø152.4
210
Unit: mm
Sample port x2 (IN, OUT) 1/4NPT
Pipe size: 1-1/2 Sch80
Approx. 690
Approx. 820
Approx.
410
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...