<9. Inspection and Maintenance>
9-14
IM 11Y01D01-01EN
9.2.4
Time Chart
The valve operation during manual online validation execution and the timing when the AO/DO
output switches to Cal/Val mode are shown below. In Cal/Val mode, it is possible to hold the AO
output or specify other settings. For the AO/DO output settings during Cal/Val mode, see “6.4.2
Output Hold” and “6.5.1 DO Contact (DO-1)”.
In Figure 9.6, Valve1 and Valve2 are switched manually by following the instructions on the
operation screen. If automatic valve control is enabled, there is no need for the operator to switch
the valves manually.
Operator
HART/HMI screen Valve1 Valve2
AO/DO
mode
[Screen operation] Start online validation 1
OFF
OFF Normal output
[Manual valve operation] Purge with check gas
Check gas purging
ON
OFF
Cal/Val
(Wait for the gas concentration to stabilize.)
[Screen operation] Proceed to the next step
[Manual valve operation] Discharge check gas
Check gas
discharging
OFF OFF
(Wait for the gas concentration to stabilize.)
[Screen operation] Touch Execute
Validation result
Normal output
Time
Figure 9.6
Valve and AO/DO output for manual online validation 1
Operator
HART/HMI screen Valve1 Valve2
AO/DO
mode
[Screen operation] Start online validation 2
OFF
OFF Normal output
[Manual valve operation] Purge with check gas
Check gas purging
ON
ON
Cal/Val
(Wait for the gas concentration to stabilize.)
[Screen operation] Proceed to the next step
[Manual valve operation] Discharge check gas
Check gas
discharging
OFF OFF
(Wait for the gas concentration to stabilize.)
[Screen operation] Touch Execute
Validation result
Normal output
Time
Figure 9.7
Valve and AO/DO output for manual online validation 2
If the initial stream is set to Stream1 in a piping arrangement with process gas switching, the
valve state transition is different as shown below. For details, see section 9.2.1.
Operator
Valve1 Valve2
[Screen operation] Start online validation 2
OFF
OFF
[Manual valve operation] Purge with check gas
OFF
ON
[Manual valve operation] Discharge check gas
OFF
OFF
9.3
Mounting on a Calibration Cell
Before performing offline validation, zero calibration, or span calibration, you usually need
to remove the LU and SCU from the process interface and mount them on a calibration cell.
However, if the TDLS8000 is mounted on a flow cell, the gas flowing through the flow cell can be
switched from the process gas making it possible to perform offline validation, zero calibration,
and span calibration without removing the TDLS8000 from the flow cell.
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...