<2. Specifications>
2-3
IM 11Y01D01-01EN
Transmitter power supply; 15 V DC or higher (at 20 mA DC)
26 V DC or less (at 0 mA DC)
Note: This voltage is generated between the AI terminals of TDLS8000. When calculating the minimum operating voltage for
transmitters, consider to allow margins for voltage drop in external wiring.
Self-diagnostics:
Laser Unit temperature, Sensor Control Unit temperature, Laser temperature, Detector
signal level, Memory read/write function, Peak locking condition
Calibration:
Calibration method; Zero/Span calibration
Calibration mode; Manual, Auto (Time initiate, Remote initiate (DI/Modbus), Semi-Auto
(YH8000/HART)
Validation:
Validation method; Up to 2 points
Validation mode;
Manual, Auto (Time initiated, Remote initiate (DI/Modbus)), Semi-Auto
(YH8000/HART)
Power supply: 24V DC +/-10%
If your power supply is 100 to 240 V AC, Universal Power Supply, M1276WW
(sold separately), is required
Power consumption:
Max. 20W; TDLS8000 only
Max. 60W; TDLS8000 with YH8000 and 2 solenoid valves
Protection degree: IP66, Type 4X
Material:
Case;
Aluminum
alloy
Wetted materials: 316 SS, BK-7 glass, Teflon encapsulated FKM (O-ring for alignment flange),
Silicone (O-ring for LAO)
Paint color: Mint green (RAL 190 30 15 or equivalent)
Weight:
Sensor Control Unit;
8 kg
Laser
Unit;
8
kg
Large Aperture Optics;
14 kg
ANSI Class 150-2-RF (Eq.) Alignment Flange;
5 kg/pc
ANSI Class 150-3-RF (Eq.) Alignment Flange;
7 kg/pc
ANSI Class 150-4-RF (Eq.) Alignment Flange; 9 kg/pc
DIN PN16-DN50-D (Eq.) Alignment Flange; 5 kg/pc
DIN PN16-DN80-D (Eq.) Alignment Flange; 6 kg/pc
JIS 10K-50-FF (Eq.) Alignment Flange;
5 kg/pc
JIS 10K-80-FF (Eq.) Alignment Flange;
6 kg/pc
Flow Cell Alignment Flange;
5 kg/pc
Process gas condition:
Process gas temperature; Max. 1,500ºC, Application dependent
Process gas pressure;
Max.1 MPa abs., Min. 90 kPa abs., Application dependent
Max. 15 kPa G with LAO unit
Note: When using TDLS8000 as CE marking compliance product, it has following limitation.
General purpose model (-G1, -G2): The upper limit of the measurement gas pressure is 50kPa in gauge pressure.
ATEX model (-S1, -S2):
The upper limit of the measurement gas pressure is 500kPa abs. The unstable gas
defined by following cannot be measured.
An unstable gas in this context is a gas liable to transform itself spontaneously, producing a sudden pressure increase.
Such transformation as an example can result from a relatively small variation of an operating parameter (e.g. pressure,
temperature, presence of catalyzing material) in a confined volume.
This includes gases that are classified as chemically unstable gases according to CLP Regulation (EC) No 1272/2008 as
amended.
Typical examples of unstable gases: acetylene (UN 1001), methyl acetylene (UN 1060), vinylfluoride (UN 1860), ozone and
dinitrogen oxide (UN 1067).
For further examples, see Table 35.1 of the UN Manual of Tests and Criteria.
Dust in process gas;
20 g/m
3
or less
(Dust loading levels are dependent upon the application, OPL
and other installation factors)
Warm-up time: 5 min.
Installation condition:
Ambient operating temperature; -20 to 55ºC
Storage temperature;
-30 to 70ºC
Humidity;
0 to 95%RH at 40ºC (Non-condensing)
Mounting flange type;
ASME B16.5, DIN, JIS
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...