03-914978-00:1
161
Miscellaneous Procedures and Instructions
3. Open the vent valve lever on the front of the mass spectrometer for ten
minutes.
4. Keep an eye on the capillary column inside the GC as you gently slide the
mass spectrometer away from the GC. Be sure not to bend or kink the
capillary column.
5. Use the alignment tool to prevent the transfer line from turning while you
loosen the brass capillary nut connecting the column to the transfer line.
6. Cap the transfer line with a capillary nut and no-hole ferrule.
7. Place the capillary column and nut inside the GC oven. This will protect
them from damage.
8. Turn off the carrier gas, then disconnect the helium gas line that is connected
to the GC filter.
9. Cap the filters with Swagelok plugs or caps.
10. Move the Saturn GC/MS to its new location. Be sure the new location
satisfies the power and environmental requirements.
Parts and Supplies
Listed below are part numbers and descriptions for the available Saturn Field
Service Parts. Items are in quantities of one (1) each unless otherwise specified.
Kits, Assemblies, Boards, and Cables
Part Number Description
03-926082-20
PWA, IEEE INTERFACE, PCI
03-930326-01 IEEE
Cable
03-930210-02
PWA, POWER BOARD
03-930250-01
PWA, RF GENERATOR
03-930300-01
Cable, SAP to PWR 26 pins (Ribbon)
03-930301-01
Cable, SAP to PWR 64 pins (Ribbon)
03-930113-91
Replacement Spares Kit
03-930010-01
Assembly, Analyzer Flange
03-930005-91
03-930005-92
Assembly, Transfer Line (115V)
Assembly, Transfer Line (230V)
03-930334-01
03-930334-02
Cable, Transfer Line heater (115V)
Cable, Transfer Line heater (230V)
03-930008-91
03-930008-92
Assembly, Vacuum Manifold (115V)
Assembly, Vacuum Manifold (230V)
03-930558-01
Assembly, Ion Gauge
03-930313-01
Cable, Trap Heater (115V)
03-930313-92
Cable, Trap Heater (230V)