03-914978-00:1
59
Chemical Ionization Options
the Manual Control comes up with the Method tab displayed. Note that the
method Name, Mode, and Scan Range are displayed.
5. Click on the Adjustments tab dialog and then choose Acetonitrile as the gas and
click on Adjust.
6. The adjustment process will begin, as shown.
7. If, after a couple of minutes, there is not enough CI gas entering the trap,
increase the flow by turning the needle valve for the chosen Channel on the MCI
Module clockwise.