03-914978-00:1
47
Chemical Ionization Options
3. Vent the mass spec.
4. Unplug the power cord from the back of the mass spec.
5. Unpack the Multiple CI kit and place the parts on a table.
6. Open the front door on the mass spec until it presses against the right side panel.
7. Find the register plate (03-930072-01) in the CI kit.
8. Peel off the adhesive strip.
9. Position the register plate 0.25 inches (0.6 cm) from the open front door and
5.25 inches (13.3 cm) from the bottom. Be sure to position the plate parallel to
the open front door. Press firmly to adhere the plate to the right side panel.
10. Close the front door and remove the right side panel.
11. Use a 5/32-inch (4 mm) drill bit to drill four holes through the side panel using the
register plate as a template.
12. There are two more holes in the register plate. Guide the metal tube through the
upper hole and the Tygon
tube through the lower hole.