2000 MS Hardware Maintenance
10
Functional Description
The Mass Spectrometer
The Saturn GC/MS employs an ultra trace ion trap mass spectrometer. The
mass spectrometer consists of the mechanical and electronic assemblies. The
following sections describe these assemblies.
The instrument is separated into the electronics and analyzer compartments.
The electronics compartment includes the
•
Scan acquisition processor/waveform board (SAP/Wave board)
•
Power board
•
The analyzer compartment includes the
•
Transfer line
•
Vacuum manifold (including the ion trap)
•
Vacuum pump and controller
•
RF coil and generator
•
Pneumatics manifold
Mechanical Assemblies
The Saturn mechanical assemblies include the following:
•
Controls and indicators
•
Cooling fans
•
Vacuum system
•
Transfer line
•
Ion trap assembly
Controls and Indicators
The main power switch for the Saturn GC/MS, located on the rear panel, controls
power to the vacuum system and to the electronics.
When the main power switch is turned on, the light-emitting diode (LED) on the
Saturn GC/MS front panel is illuminated.
In the event of an emergency, shut off of all power to the Saturn GC/MS by
placing the main power switch in the OFF position and unplugging the
instrument.