GC Maintenance
Refer to the figure below when removing and/or replacing the glass insert.
1 Injector
Nut
2 Septum
3 Septum
Purge
4 Septum
Support
5 Graphite
Ferrule
6
Carrier Gas Inlet
7 Split
Vent
8 Glass
Insert
9
Fused Silica Column
10 Capillary Column Nut
11 Vespel®
Ferrule
12 Point of Injection
Tools Required
Tweezers or septum pick
(72-000084-00)
Injector nut wrench (03-908423-00)
Flat-blade screwdriver (short handle)
Clean laboratory tissue
Graphite ferrules (03-925342-02)
Insert/ferrule positioning tool
(03-925385-00)
1079 Injector - Cross-Sectional View
Remove the Glass Insert
Follow these steps to remove the glass insert from the 1079 Injector:
1. Use the injector nut wrench to remove the injector nut (Item 1). Place the nut
on a clean surface (e.g., clean tissue).
WARNING:
BURN HAZARD
The injector nut may be hot. Lower the injector temperature to 50
°°°°
C and permit
the injector nut to cool before proceeding.