Introduction
Page 1-4
_________________________________________________________________
Model
2010D
Gas Dilution Calibrator Specifications
Pneumatics:
Flow ac 1.0% of full scale
Flow repeata 0.15% of full scale
Flow lin 0.5% of full scale
Diluent:
Mass flow controller range 0 - 10 SLPM (Optional 0 - 20 SLPM)
Input pressure 20 - 30 PSIG
Port 1, normally air (Optional Port 2, user defined)
Source:
Mass flow controller range 0 - 100 SCCM (Optional 0 - 20 SCCM,
0 - 50 SCCM, 0 - 200 SCCM, 0 - 500 SCCM, 0 - 1 SLPM, 0 - 2 SLPM)
Optional second source mass flow controller
Input pressure 15 - 30 PSIG
Four gas cylinder input ports (Optional two additional ports)
Source manifold purge port
Optional external permeation oven input port
Usable dilution ratio varies with flow controller selection, add
a second source flow controller for wider dynamic range
Solenoid valves for diluent, source and purge ports
Eight calibration gas output ports
Optional output solenoid valve
GPT reaction chamber meets EPA requirements
Attains 95% of Set Point in less than 60 Seconds
Wetted surfaces are composed of stainless steel, Teflon, Peek,
Halar or glass; all ports Swagelok or equivalent; Viton and Calrez
seals
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......